FFLS: An accurate linear device for measuring synergistic finger contractions
Kõiva R, Hilsenbeck B, Castellini C (2012)
Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA.
Konferenzbeitrag
| Veröffentlicht | Englisch
Download
2012_Koiva_Hilsenbeck_Castellini__FFLS_An_accurate_linear_device_for_measuring.pdf
Autor*in
Kõiva, RistoUniBi ;
Hilsenbeck, Barbara;
Castellini, Claudio
Einrichtung
Abstract / Bemerkung
After decades of theoretical study in physiology and neurology communities, the paradigm of muscle synergies is now being explored in rehabilitation robotics as a strategy to control mechanical artifacts with many degrees-of-freedom (DoF) in a simple yet effective and human-like way. In particular, muscle synergies during grasping and in graded-force tasks are of great interest for the control of dexterous hand prostheses.
To this end, we have designed and tested a novel device to accurately and simultaneously measure fingertip forces. The device, called FFLS (Finger-Force Linear Sensor), measures the forces applied by the human fingertips in both directions (flexion and extension of index, middle, ring and little finger plus thumb rotation and abduction/adduction). It is suited for several different hand sizes, enforces high accuracy in the measurement and its signal is guaranteed to be linear in a high range of forces (100N in both directions for each finger). It outputs six analog voltages (+/-10V), suited for processing with a DAQ card.
Stichworte
Hand force measurement
Erscheinungsjahr
2012
Konferenz
34th Annual International IEEE EMBS Conference (EMBC 2012)
Konferenzort
San Diego, USA
Konferenzdatum
2012-08-29
ISBN
978-1-4577-1787-1
Page URI
https://pub.uni-bielefeld.de/record/2485135
Zitieren
Kõiva R, Hilsenbeck B, Castellini C. FFLS: An accurate linear device for measuring synergistic finger contractions. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA.
Kõiva, R., Hilsenbeck, B., & Castellini, C. (2012). FFLS: An accurate linear device for measuring synergistic finger contractions. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA. doi:10.1109/embc.2012.6345985
Kõiva, Risto, Hilsenbeck, Barbara, and Castellini, Claudio. 2012. “FFLS: An accurate linear device for measuring synergistic finger contractions”. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA .
Kõiva, R., Hilsenbeck, B., and Castellini, C. (2012).“FFLS: An accurate linear device for measuring synergistic finger contractions”. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA.
Kõiva, R., Hilsenbeck, B., & Castellini, C., 2012. FFLS: An accurate linear device for measuring synergistic finger contractions. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA.
R. Kõiva, B. Hilsenbeck, and C. Castellini, “FFLS: An accurate linear device for measuring synergistic finger contractions”, Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA, 2012.
Kõiva, R., Hilsenbeck, B., Castellini, C.: FFLS: An accurate linear device for measuring synergistic finger contractions. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA (2012).
Kõiva, Risto, Hilsenbeck, Barbara, and Castellini, Claudio. “FFLS: An accurate linear device for measuring synergistic finger contractions”. Presented at the 34th Annual International IEEE EMBS Conference (EMBC 2012), San Diego, USA, 2012.
Volltext(e)
Name
2012_Koiva_Hilsenbeck_Castellini__FFLS_An_accurate_linear_device_for_measuring.pdf
Access Level
UniBi Only
Zuletzt Hochgeladen
2019-09-06T09:18:00Z
MD5 Prüfsumme
c6e56ae9411e27cd938ea257d8c77300