Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass

Hassanein W, Albert AA, Arnrich B, Walter JA, Ennker IC, Rosendahl UP, Bauer S, Ennker J (2005)
The annals of thoracic surgery 80(6): 2155-2161.

Zeitschriftenaufsatz | Veröffentlicht | Englisch
 
Download
Es wurden keine Dateien hochgeladen. Nur Publikationsnachweis!
Autor*in
Hassanein, Wael; Albert, Alexander A.; Arnrich, Bert; Walter, Jörg A.; Ennker, Ina Carolin; Rosendahl, Ulrich P.; Bauer, Stefan; Ennker, Jürgen
Abstract / Bemerkung
Background Off-pump coronary artery bypass grafting (OPCAB) has attracted increasing attention. Performing the anastomosis off-pump is technically more demanding. The objective of the study is to assess the quality of anastomosis in OPCAB in comparison with conventional on-pump coronary artery bypass grafting using the transit time flow measurement. Methods Four hundred forty-five patients operated on using OPCAB technique were included in the study. For each patient in this group a similar patient from the on-pump coronary artery bypass grafting population was selected according to the number of grafts, bypass material, and target coronary arteries. The mean flow and the pulsatile index were measured in every bypass graft in both groups. Results The average pulsatile index in OPCAB was 2.09 ± 1.03 (mean flow, 39 ± 22.63 mL/min), whereas with on-pump coronary artery bypass grafting it was 1.9 ± 0.98 (mean flow, 44.19 ± 23.58 mL/min); p = 0.005. Subgroup analysis showed significantly lower mean flows and higher pulsatile index with OPCAB in grafts to the obtuse marginal, diagonal, and right coronary artery, but not to the left anterior descending territory. Conclusions The quality of the anastomosis performed using the OPCAB technique might be jeopardized by less accessibility as in the case of lateral and posterior wall coronary arteries. Techniques to optimize the accessibility of the coronary artery like combining sling support with cup stabilizers, together with systematic training, should be strongly considered in OPCAB. Whenever there is good accessibility of the coronary artery as in the case of the left anterior descending, the anastomosis performed under OPCAB has a quality as good as that performed using the conventional technique.
Erscheinungsjahr
2005
Zeitschriftentitel
The annals of thoracic surgery
Band
80
Ausgabe
6
Seite(n)
2155-2161
ISSN
0003-4975
Page URI
https://pub.uni-bielefeld.de/record/2714299

Zitieren

Hassanein W, Albert AA, Arnrich B, et al. Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass. The annals of thoracic surgery. 2005;80(6):2155-2161.
Hassanein, W., Albert, A. A., Arnrich, B., Walter, J. A., Ennker, I. C., Rosendahl, U. P., Bauer, S., et al. (2005). Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass. The annals of thoracic surgery, 80(6), 2155-2161. doi:10.1016/j.athoracsur.2005.03.138
Hassanein, Wael, Albert, Alexander A., Arnrich, Bert, Walter, Jörg A., Ennker, Ina Carolin, Rosendahl, Ulrich P., Bauer, Stefan, and Ennker, Jürgen. 2005. “Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass”. The annals of thoracic surgery 80 (6): 2155-2161.
Hassanein, W., Albert, A. A., Arnrich, B., Walter, J. A., Ennker, I. C., Rosendahl, U. P., Bauer, S., and Ennker, J. (2005). Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass. The annals of thoracic surgery 80, 2155-2161.
Hassanein, W., et al., 2005. Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass. The annals of thoracic surgery, 80(6), p 2155-2161.
W. Hassanein, et al., “Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass”, The annals of thoracic surgery, vol. 80, 2005, pp. 2155-2161.
Hassanein, W., Albert, A.A., Arnrich, B., Walter, J.A., Ennker, I.C., Rosendahl, U.P., Bauer, S., Ennker, J.: Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass. The annals of thoracic surgery. 80, 2155-2161 (2005).
Hassanein, Wael, Albert, Alexander A., Arnrich, Bert, Walter, Jörg A., Ennker, Ina Carolin, Rosendahl, Ulrich P., Bauer, Stefan, and Ennker, Jürgen. “Intraoperative transit time flow measurement: off-pump versus on-pump coronary artery bypass”. The annals of thoracic surgery 80.6 (2005): 2155-2161.

13 Zitationen in Europe PMC

Daten bereitgestellt von Europe PubMed Central.

Influence of coronary territory on flow profiles of saphenous vein grafts.
Amin S, Werner RS, Madsen PL, Krasopoulos G, Taggart DP., J Cardiothorac Surg 13(1), 2018
PMID: 29463268
Intraoperative graft assessment during coronary artery bypass surgery.
Fukui T., Gen Thorac Cardiovasc Surg 63(3), 2015
PMID: 25556862
New device for intraoperative graft assessment: HyperEye charge-coupled device camera system.
Handa T, Katare RG, Nishimori H, Wariishi S, Fukutomi T, Yamamoto M, Sasaguri S, Sato T., Gen Thorac Cardiovasc Surg 58(2), 2010
PMID: 20155342
Intraoperative grafts assessment.
Leacche M, Balaguer JM, Byrne JG., Semin Thorac Cardiovasc Surg 21(3), 2009
PMID: 19942118
The effects of on-pump and off-pump coronary artery bypass grafting on intraoperative graft flow in arterial and venous conduits defined by a flow/pressure ratio.
Balacumaraswami L, Abu-Omar Y, Selvanayagam J, Pigott D, Taggart DP., J Thorac Cardiovasc Surg 135(3), 2008
PMID: 18329465
Intraoperative imaging techniques to assess coronary artery bypass graft patency.
Balacumaraswami L, Taggart DP., Ann Thorac Surg 83(6), 2007
PMID: 17532450
Measurement of the flow in coronary artery bypass grafts.
Nemec P, Bruk V, Steriovsky A, Gwozdzeiwicz M, Hajek R., Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub 150(1), 2006
PMID: 16936916
Graft-to-coronary-artery shunt during off-pump coronary artery bypass grafting.
Iida H, Mori H, Sudo Y, Yamada Y, Eda K, Inoue Y., Ann Thorac Surg 82(3), 2006
PMID: 16928480
Invited commentary.
Mihaljevic T., Ann Thorac Surg 80(6), 2005
PMID: 16305863

28 References

Daten bereitgestellt von Europe PubMed Central.

Coronary bypass graft fate and patient outcome: angiographic follow-up of 5,065 grafts related to survival and reoperation in 1,388 patients during 25 years.
Fitzgibbon GM, Kafka HP, Leach AJ, Keon WJ, Hooper GD, Burton JR., J. Am. Coll. Cardiol. 28(3), 1996
PMID: 8772748
Does multivessel, off-pump coronary artery bypass reduce postoperative morbidity?
Kshettry VR, Flavin TF, Emery RW, Nicoloff DM, Arom KV, Petersen RJ., Ann. Thorac. Surg. 69(6), 2000
PMID: 10892915
Cytokine responses and myocardial injury in coronary artery bypass grafting.
Wei M, Kuukasjarvi P, Laurikka J, Kaukinen S, Iisalo P, Laine S, Laippala P, Metsanoja R, Tarkka M., Scand. J. Clin. Lab. Invest. 61(2), 2001
PMID: 11347983
Graft revision after transit time flow measurement in off-pump coronary artery bypass grafting.
D'Ancona G, Karamanoukian HL, Ricci M, Schmid S, Bergsland J, Salerno TA., Eur J Cardiothorac Surg 17(3), 2000
PMID: 10758390
Clinical importance of measuring coronary graft flows in the revascularized heart. Ultrasonic or electromagnetic?
Canver CC, Cooler SD, Murray EL, Nichols RD, Heisey DM., J Cardiovasc Surg (Torino) 38(3), 1997
PMID: 9219468
Data mart based research in heart surgery
Arnrich, Medinfo 11(), 2004
"Single suture" for circumflex exposure in off-pump coronary artery bypass grafting.
Bergsland J, Karamanoukian HL, Soltoski PR, Salerno TA., Ann. Thorac. Surg. 68(4), 1999
PMID: 10543532
Off-pump coronary artery bypass surgery: physiology and anaesthetic management.
Chassot PG, van der Linden P, Zaugg M, Mueller XM, Spahn DR., Br J Anaesth 92(3), 2004
PMID: 14970136
Mammary artery-coronary artery anastomosis as method of treatment for angina pectoris.
Kolessov VI., J. Thorac. Cardiovasc. Surg. 54(4), 1967
PMID: 6051440
Complement and the damaging effects of cardiopulmonary bypass.
Kirklin JK, Westaby S, Blackstone EH, Kirklin JW, Chenoweth DE, Pacifico AD., J. Thorac. Cardiovasc. Surg. 86(6), 1983
PMID: 6606084
The factors effecting complement activation in open heart surgery.
Aslan R, Tunerir B, Dernek S, Kural T, Sevin B, Kaklikkaya I, Manti Z., J Cardiovasc Surg (Torino) 33(6), 1992
PMID: 1287017
Biological effects of off-pump vs. on-pump coronary artery surgery: focus on inflammation, hemostasis and oxidative stress.
Biglioli P, Cannata A, Alamanni F, Naliato M, Porqueddu M, Zanobini M, Tremoli E, Parolari A., Eur J Cardiothorac Surg 24(2), 2003
PMID: 12895618
Validation of a new transit time ultrasound flowmeter in man.
Laustsen J, Pedersen EM, Terp K, Steinbruchel D, Kure HH, Paulsen PK, Jorgensen H, Paaske WP., Eur J Vasc Endovasc Surg 12(1), 1996
PMID: 8696905
Evaluation of coronary bypass flow with color-Doppler and magnetic resonance imaging techniques: comparison with intraoperative flow measurements.
Walpoth BH, Muller MF, Genyk I, Aeschbacher B, Kipfer B, Althaus U, Carrel TP., Eur J Cardiothorac Surg 15(6), 1999
PMID: 10431861
Transit-time flow measurement for detection of early graft failure during myocardial revascularization.
Walpoth BH, Bosshard A, Genyk I, Kipfer B, Berdat PA, Hess OM, Althaus U, Carrel TP., Ann. Thorac. Surg. 66(3), 1998
PMID: 9769011
Flow measurement in coronary surgery.
D'Ancona G, Karamanoukian HL, Salerno TA, Schmid S, Bergsland J., Heart Surg Forum 2(2), 1999
PMID: 11276468
Transit time flow measurement in off-pump coronary artery bypass grafting
D’Ancona, 2001
An evaluation of the intraoperative transit time measurements of coronary bypass flow.
Hirotani T, Kameda T, Shirota S, Nakao Y., Eur J Cardiothorac Surg 19(6), 2001
PMID: 11404141
Can visual assessment of flow waveform morphology detect anastomotic error in off-pump coronary artery bypass grafting?
Jaber SF, Koenig SC, BhaskerRao B, VanHimbergen DJ, Spence PA., Eur J Cardiothorac Surg 14(5), 1998
PMID: 9860203
Transit time flow measurement in on-pump and off-pump coronary artery surgery.
Schmitz C, Ashraf O, Schiller W, Preusse CJ, Esmailzadeh B, Likungu JA, Fimmers R, Welz A., J. Thorac. Cardiovasc. Surg. 126(3), 2003
PMID: 14502134
Off-pump coronary artery bypass may decrease the patency of saphenous vein grafts.
Kim KB, Lim C, Lee C, Chae IH, Oh BH, Lee MM, Park YB., Ann. Thorac. Surg. 72(3), 2001
PMID: 11565721
A randomized comparison of off-pump and on-pump multivessel coronary-artery bypass surgery.
Khan NE, De Souza A, Mister R, Flather M, Clague J, Davies S, Collins P, Wang D, Sigwart U, Pepper J., N. Engl. J. Med. 350(1), 2004
PMID: 14702424
Clinical outcomes, angiographic patency, and resource utilization in 200 consecutive off-pump coronary bypass patients.
Puskas JD, Thourani VH, Marshall JJ, Dempsey SJ, Steiner MA, Sammons BH, Brown WM 3rd, Gott JP, Weintraub WS, Guyton RA., Ann. Thorac. Surg. 71(5), 2001
PMID: 11383786
Coronary artery bypass grafting with cardiopulmonary bypass versus off-pump cardiopulmonary bypass grafting: does eliminating the pump reduce morbidity and cost?
Bull DA, Neumayer LA, Stringham JC, Meldrum P, Affleck DG, Karwande SV., Ann. Thorac. Surg. 71(1), 2001
PMID: 11216740
Seven-year follow-up of coronary artery bypasses performed with and without cardiopulmonary bypass.
Gundry SR, Romano MA, Shattuck OH, Razzouk AJ, Bailey LL., J. Thorac. Cardiovasc. Surg. 115(6), 1998
PMID: 9628668

AUTHOR UNKNOWN, 0
Export

Markieren/ Markierung löschen
Markierte Publikationen

Open Data PUB

Web of Science

Dieser Datensatz im Web of Science®
Quellen

PMID: 16305862
PubMed | Europe PMC

Suchen in

Google Scholar