Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees

Eichler T, Ibrahim A, Riemann C, Scholtz LU, Gehl HB, Goon P, Sudhoff H, Todt I (2022)
Brain Sciences 12(10): 1406.

Zeitschriftenaufsatz | Veröffentlicht | Englisch
 
Download
OA 471.53 KB
Autor*in
Eichler, Theda; Ibrahim, Ahmed; Riemann, Conrad; Scholtz, Lars Uwe; Gehl, Hans Björn; Goon, PeterUniBi; Sudhoff, HolgerUniBi ; Todt, IngoUniBi
Abstract / Bemerkung
Introduction: The approval process for MRI safety of implants includes physical observations and an experimental evaluation in artificial settings to simulate the in vivo effect. This contains the observation of temperature changes and artificial current generation by the magnetic field. From these findings, the safety of an implant and its effect on the patient can be estimated. MRI safety is based on an in vivo evaluation of adverse events after the approval process, but an actual analysis of the effect on different tissues is not followed. The effect of MRI scanning in cochlea implantees on their residual hearing as the correlate of the hair cell function is so far unknown, therefore the aim of the present study was to observe the effect of 3 T MRI on the residual hearing of cochlea implantees. Material and Methods: In this prospective study, we performed a 3 T MRI T2 2D MS Drive sequence in eight cochlea-implanted ears. Before and after the MRI scan, a bone conduction pure tone audiogram (BC PTA) was performed. All cochlea implantees had a pre-scanning threshold of low frequency residual hearing between 20 dB and 65 dB. Results: Low frequency mean residual hearing was not affected by the 3 T T2 2D MS Drive sequence. We observed a pre-scanning threshold at 250 Hz of 42.9 (SD 3.9) dB and for 500 Hz 57.1 (SD 6.4) dB. Post-scanning BC PTA was for 250 Hz 42.1 (SD 3.9) dB and for 500 Hz 57.1 (SD 5.7) dB. Conclusion: 3 T MRI scanning has no significant functional effect on the hair cells in cochlea implantees in low frequencies with a T2 2D MS Drive sequence.
Erscheinungsjahr
2022
Zeitschriftentitel
Brain Sciences
Band
12
Ausgabe
10
Art.-Nr.
1406
eISSN
2076-3425
Page URI
https://pub.uni-bielefeld.de/record/2966427

Zitieren

Eichler T, Ibrahim A, Riemann C, et al. Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees. Brain Sciences. 2022;12(10): 1406.
Eichler, T., Ibrahim, A., Riemann, C., Scholtz, L. U., Gehl, H. B., Goon, P., Sudhoff, H., et al. (2022). Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees. Brain Sciences, 12(10), 1406. https://doi.org/10.3390/brainsci12101406
Eichler, Theda, Ibrahim, Ahmed, Riemann, Conrad, Scholtz, Lars Uwe, Gehl, Hans Björn, Goon, Peter, Sudhoff, Holger, and Todt, Ingo. 2022. “Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees”. Brain Sciences 12 (10): 1406.
Eichler, T., Ibrahim, A., Riemann, C., Scholtz, L. U., Gehl, H. B., Goon, P., Sudhoff, H., and Todt, I. (2022). Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees. Brain Sciences 12:1406.
Eichler, T., et al., 2022. Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees. Brain Sciences, 12(10): 1406.
T. Eichler, et al., “Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees”, Brain Sciences, vol. 12, 2022, : 1406.
Eichler, T., Ibrahim, A., Riemann, C., Scholtz, L.U., Gehl, H.B., Goon, P., Sudhoff, H., Todt, I.: Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees. Brain Sciences. 12, : 1406 (2022).
Eichler, Theda, Ibrahim, Ahmed, Riemann, Conrad, Scholtz, Lars Uwe, Gehl, Hans Björn, Goon, Peter, Sudhoff, Holger, and Todt, Ingo. “Prospective Evaluation of 3 T MRI Effect on Residual Hearing Function of Cochlea Implantees”. Brain Sciences 12.10 (2022): 1406.
Alle Dateien verfügbar unter der/den folgenden Lizenz(en):
Creative Commons Namensnennung 4.0 International Public License (CC-BY 4.0):
Volltext(e)
Access Level
OA Open Access
Zuletzt Hochgeladen
2022-10-26T07:44:06Z
MD5 Prüfsumme
280e6452aeb4ee548a25d77072ef3c0e


Export

Markieren/ Markierung löschen
Markierte Publikationen

Open Data PUB

Web of Science

Dieser Datensatz im Web of Science®
Quellen

PMID: 36291339
PubMed | Europe PMC

Suchen in

Google Scholar